Enviro-T2™ In-Line Fluorometer
Ideal for water treatment facilities wanting to continuously monitor levels of algae or cyanobacteria, Enviro-T2™ is an accurate, single-channel fluorometer that easily installs in-line and integrates with data collection systems. When configured with a Blue Excitation LED, Enviro-T2 detects fluorescence from all algal groups; when configured with a Red Excitation LED, it has maximized sensitivity for better detection of Cyanobacteria.
Enviro-T2's durable and rugged flow-through system operates in both marine and freshwater conditions. Simply connect the analog output cable to a SCADA system, data logger or any control cabinet that accepts a 4-20 mA output. The wide variety of control cabinets available allows users to choose from many different features and manufacturers.
Enviro-T2 uses solid-state optoelectronics making it a reliable, high performance, rugged continuous sampling device. Instrument maintenance or inspection is very easy with a quick one-step disconnect. Factory-adjusted, Enviro-T2 requires minimal installation effort. Standard solutions are available for quick functional checks.
Patent Number 7,099,012
Product Highlights:
· Estimate Algal Abundance
- Algae
- Cyanobacteria
· Low Limit of Detection
· Easy Integration with Controllers
· Minimal Maintenance
· Compact Package
Enviro-T2 In-Line Fluorometer Specifications
Enviro-T2 Performance Specifications
Linearity: 0.99R2
Application
Minimum Detection Limit
Dynamic Range
Chlorophyll
0.03 _6;g/L
0-100 _6;g/L
Enviro-T2 Physical Specifications
Material
PVC – rigid injection-molded plastic
Weight
4.4 ounces (125 g)
Length
8.63 inches (21.9 cm)
Housing Diameter
1.3 inches (3.3 cm)
Diameter at Flange
1.72 inches (4.4 cm)
Enviro-T2 Electrical Specifications
Power Draw
0.8 W @ 12 VDC (1.0 W maximum)
Input Voltage
8-30 VDC
Signal Output
4-20 mA
Warm-up time
5 Seconds
Mounting Tee Physical Specifications
Material
PVC – rigid injection-molded plastic
Dynamic Pressure
100 PSI
Threading
1 inch (NPT)
Length